10.1007/s40204-016-0056-4

Physical and mechanical characterization of PLLA interference screws produced by two stage injection molding method

  1. Department of Materials Research, Iranian Academic Center for Education, Culture and Research (ACECR), Mashhad Branch, Mashhad, IR
  2. Department of Materials Research, Iranian Academic Center for Education, Culture and Research (ACECR), Mashhad Branch, Mashhad, IR Materials Engineering Department, Tarbiat Modares University, Tehran, IR
  3. Department of Mechanical Engineering, Tarbiat Modares University, Tehran, IR
Cover Image

Published in Issue 2016-10-22

How to Cite

Sadeghi-Avalshahr, A. R., Khorsand-Ghayeni, M., Nokhasteh, S., Molavi, A. M., & Sadeghi-Avalshahr, M. (2016). Physical and mechanical characterization of PLLA interference screws produced by two stage injection molding method. Progress in Biomaterials, 5(3-4 (December 2016). https://doi.org/10.1007/s40204-016-0056-4

Abstract

Abstract The purpose of this study was to produce and evaluate different mechanical, physical and in vitro cell culture characteristics of poly(L-lactic) acid (PLLA) interference screws. This work will focus on evaluating the effect of two important parameters on operation of these screws, first the tunnel diameter which is one of the most important parameters during the operation and second the thermal behavior, the main effective characteristic in production process. In this work, PLLA screws were produced by a two-stage injection molding machine. For mechanical assessment of the produced screws, Polyurethane rigid foam was used as cancellous bone and polypropylene rope as synthetic graft to simulate bone and ligament in real situation. Different tunnel diameters including 7–10 mm were evaluated for fixation strength. When the tunnel diameter was changed from 10 to 9 mm, the pull-out force has increased to about 12 %, which is probably due to the aforementioned frictional forces, however, by reducing the tunnel diameter to 8 and 7 mm, the pull-out force reduced to 16 and 50 % for 8 and 7 mm tunnel diameter, respectively. The minimum and maximum pull-out force was obtained 160.57 and 506.86 N for 7 and 9 mm tunnel diameters, respectively. For physicochemical assay, Fourier transform infrared spectroscopy (FTIR), degradation test and differential scanning calorimetry (DSC) were carried out. The crystallinity (Xc) of samples were decreased considerably from 64.3 % before injection to 32.95 % after injection with two different crystallographic forms α′ and α. probably due to the fast cooling rate at room temperature. In addition, MTT and cell attachment assays were utilized by MG63 osteoblast cell line, to evaluate the cytotoxicity of the produced screws. The results revealed no cytotoxicity effect.

Keywords

  • Interference screw,
  • Biodegradable,
  • PLLA,
  • ACL reconstruction

References

  1. Androsch et al. (2014) Melting of conformationally disordered crystals (α′-Phase) of Poly(L-lactic acid) (pp. 1134-1139) https://doi.org/10.1002/macp.201400126
  2. Armentano et al. (2010) Biodegradable polymer matrix nanocomposites for tissue engineering: a review (pp. 2126-2146) https://doi.org/10.1016/j.polymdegradstab.2010.06.007
  3. Bailey et al. (2006) Biomechanical evaluation of a new composite bioresorbable screw (pp. 208-212) https://doi.org/10.1016/j.jhsb.2005.10.015
  4. Barber (1999) Tripled semitendinosus-cancellous bone anterior cruciate ligament reconstruction with bioscrew fixation (pp. 360-367) https://doi.org/10.1016/S0749-8063(99)70052-6
  5. Barber (2005) Poly-D, L-Lactide interference screws for anterior cruciate ligament reconstruction (pp. 804-808) https://doi.org/10.1016/j.arthro.2005.04.104
  6. Battula et al. (2008) The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone (pp. 990-995) https://doi.org/10.1097/TA.0b013e31802bf051
  7. Carrasco et al. (2010) Processing of poly(lactic acid): characterization of chemical structure, thermal stability and mechanical properties (pp. 116-125) https://doi.org/10.1016/j.polymdegradstab.2009.11.045
  8. Chieng et al. (2013) Poly(lactic acid)/Poly(ethylene glycol) polymer nanocomposites: effects of graphene nanoplatelets (pp. 93-104) https://doi.org/10.3390/polym6010093
  9. Dhillon et al. (2016) Drill bit failure and retrieval during arthroscopic ACL reconstruction https://doi.org/10.1016/j.jcot.2016.06.008
  10. Díaz-Gómez L, Concheiro A, Alvarez-Lorenzo C (2015) Polymers in drug delivery: fundamentals. In: Puoci F (ed) Advanced polymers in medicine. Springer, Cham, pp 319–339
  11. Gleadall and Pan (2015) Mechanical properties of biodegradable polymers for medical applications (pp. 163-199) Elsevier https://doi.org/10.1533/9781782420255.2.163
  12. Gorrasi and Pantani (2013) Effect of PLA grades and morphologies on hydrolytic degradation at composting temperature: assessment of structural modification and kinetic parameters (pp. 1006-1014) https://doi.org/10.1016/j.polymdegradstab.2013.02.005
  13. Hunt and Callaghan (2008) Polymer-hydroxyapatite composite versus polymer interference screws in anterior cruciate ligament reconstruction in a large animal model (pp. 655-660) https://doi.org/10.1007/s00167-008-0528-8
  14. ISO 10993 (2009) Biological evaluation of medical devices. Part 5: tests for in vitro cytotoxicity. Int Organ Stand 5:2009
  15. Jamshidian et al. (2010) Poly-Lactic Acid: production, applications, nanocomposites, and release studies (pp. 552-571) https://doi.org/10.1111/j.1541-4337.2010.00126.x
  16. Kaeding et al. (2005) A prospective randomized comparison of bioabsorbable and titanium anterior cruciate ligament interference screws (pp. 147-151) https://doi.org/10.1016/j.arthro.2004.09.012
  17. Konan and Haddad (2009) The unpredictable material properties of bioabsorbable PLC interference screws and their adverse effects in ACL reconstruction surgery (pp. 293-297) https://doi.org/10.1007/s00167-008-0684-x
  18. Kousa et al. (2001) Initial fixation strength of bioabsorbable and titanium interference screws in anterior cruciate ligament reconstruction. Biomechanical evaluation by single cycle and cyclic loading (pp. 420-425)
  19. Kurzweil et al. (1995) Tibial interference screw removal following anterior cruciate ligament reconstruction (pp. 289-291) https://doi.org/10.1016/0749-8063(95)90004-7
  20. Lambert (1983) Vascularized patellar tendon graft with rigid internal fixation for anterior cruciate ligament insufficiency (pp. 85-89) https://doi.org/10.1097/00003086-198301000-00016
  21. Lim et al. (2008) Processing technologies for poly(lactic acid) (pp. 820-852) https://doi.org/10.1016/j.progpolymsci.2008.05.004
  22. Menard KP (2008) Dynamic mechanical analysis: a practical introduction, 2nd edn. CRC Press, Boca Raton, FL
  23. Noyes et al. (1984) Biomechanical analysis of human knee ligament grafts used in knee ligament repairs and reconstructions (pp. 344-352) https://doi.org/10.2106/00004623-198466030-00005
  24. Pan et al. (2008) Polymorphic transition in disordered poly(L-lactide) crystals induced by annealing at elevated temperatures (pp. 4296-4304) https://doi.org/10.1021/ma800343g
  25. Pantani and Sorrentino (2013) Influence of crystallinity on the biodegradation rate of injection-moulded poly(lactic acid) samples in controlled composting conditions (pp. 1089-1096) https://doi.org/10.1016/j.polymdegradstab.2013.01.005
  26. Prodromos C, Brown C, Fu FH, et al (2007) The anterior cruciate ligament: reconstruction and basic science. Elsevier Health Sciences
  27. Rasselet et al. (2014) Oxidative degradation of polylactide (PLA) and its effects on physical and mechanical properties (pp. 109-116) https://doi.org/10.1016/j.eurpolymj.2013.10.011
  28. Saeidlou et al. (2012) Poly(lactic acid) crystallization (pp. 1657-1677) https://doi.org/10.1016/j.progpolymsci.2012.07.005
  29. Shasteen and Choy (2011) Controlling degradation rate of poly(lactic acid) for its biomedical applications (pp. 163-167) https://doi.org/10.1007/s13534-011-0025-8
  30. Sidhu and Wroble (1997) Intraarticular migration of a femoral interference fit screw: a complication of anterior cruciate ligament reconstruction (pp. 268-271) https://doi.org/10.1177/036354659702500223
  31. Sultana (2011) Study of in vitro degradation of biodegradable polymer based thin films and tissue engineering scaffolds (pp. 18709-18715) https://doi.org/10.5897/AJB11.2742
  32. Weiler et al. (2000) The influence of screw geometry on hamstring tendon interference fit fixation (pp. 356-359)
  33. Wunderlich (2005) Springer Berlin
  34. Zhang et al. (2008) Disorder-to-order phase transition and multiple melting behavior of Poly(L-lactide) investigated by simultaneous measurements of WAXD and DSC (pp. 1352-1357) https://doi.org/10.1021/ma0706071